PRA number
2018/17121
Published date
Product description
Nokia Blood Pressure Monitor Plus (BPM+)
Model WPM03
ARTG #223033
Model WPM03
ARTG #223033
Reasons the product is recalled
The blood pressure monitor did not pass applicable testing standards for accuracy for subjects with an arm circumference greater than 35cm.
The hazards to consumers
The user may be unable to obtain an accurate reading of the blood pressure, including systolic and diastolic readings, if the subject's arm circumference is greater than 35cm.
What consumers should do
Users should have been contacted by an email in June 2018 and an application update in July 2018, which requires them to select one of the following options:
1. Keep their Nokia BPM+, which requires that the user acknowledges that they have been notified
2. Return their Nokia BPM+ and receive a refund. If this option is selected a link to the return form is provided.
For more information, contact Emergo Customer Care on 02 9006 1662.
1. Keep their Nokia BPM+, which requires that the user acknowledges that they have been notified
2. Return their Nokia BPM+ and receive a refund. If this option is selected a link to the return form is provided.
For more information, contact Emergo Customer Care on 02 9006 1662.
Where the product was sold
Nationally
Responsible regulator
Therapeutic Goods Administration is the responsible regulator for this recall.