Fluke—170 Series Digital Multimeter

What are the defects?

When measuring ac voltage above 500, the display on affected meters can be slow in showing a final, accurate measurement. The final reading may not be displayed for 6 to 18 seconds. During this delay, the meter may not indicate that high voltage is present.

What are the hazards?

Electric shock.

What should consumers do?

Stop using the meter and contact VMS Laboratory on Phone (02) 9771 9300 or Facsimile (02) 9771 9600 (Attention Fluke 170 recall administrator) to arrange for repairs free of charge.

Supplier
Fluke Australia Pty Ltd
Where the product was sold
Nationally

Responsible regulator

Australian Competition and Consumer Commission is the responsible regulator for this recall.