Fluke—T2 Electrical Tester

PRA number

1998/3706

Published date

Product description

Serial Numbers Less Than 70524051. Units That Have An "r" Stamped After The Serial Number Have Been Repaired And Are No Longer Affected. Sold In Australia Since February 1998.

Reasons the product is recalled

A Battery Contact To The Circuit Board Inside The Instrument Can Corrode Over Time Causing It To Operate Intermittently. When The Malfunction Occurs The Meter May Not Indicate That Voltage Is Present Placing The User In Potentially Hazardous Situation.

The hazards to consumers

Electric Shock

What consumers should do

For Information Ring Philips Business Electronics On (02) 9805 4486. Units Will Be Serviced And Returned Directly To Customers At No Charge.

Supplier

Fluke Australia Pty Ltd

Where the product was sold

Nationally

Responsible regulator

Australian Competition and Consumer Commission is the responsible regulator for this recall.

Product category

Is this page useful?