GE Healthcare Australia Pty Ltd—GE Healthcare TruSat Pulse Oximeter

Identifying features

Reference numbers:
6051-0000-190,
6051-0000-191,
6051-0000-192 and
6051-0000-193

ARTG number 139421

What are the defects?

The external power supply does not conform with current safety standards. It is possible that the external power supply could leak electrical current.

What are the hazards?

Potentially life threatening or could cause a serious health risk (Class I)

What should consumers do?

Home care patients will be contacted individually by their treating doctor.

Supplier
GE Healthcare Australia Pty Ltd
Where the product was sold
New South Wales
Northern Territory
Queensland
Victoria
Western Australia

Responsible regulator

Therapeutic Goods Administration is the responsible regulator for this recall.