Product description

Accu-Chek Solo Cannula Assembly & Pump Holder (9 mm; 10+3 & 6 mm; 10+3) and Accu-Chek Solo Supply Kit (9 mm) & (6 mm)

Cannula lot numbers: 847468303 and lower

Reference numbers: 07858833001, 07858841001, 07858817001 & 07858825001

ARTG 308704
(Roche Diabetes Care Australia Pty Ltd - Insulin infusion patch pump administration set)

ARTG 311432
(Roche Diabetes Care Australia Pty Ltd - Ambulatory insulin infusion administration set/reservoir kit)

See a list of details to help identify the product

Reason the product is recalled

Insulin leakage has occurred with the Accu-Chek Solo Cannula Assembly and  Pump Holder.

If the insulin leakage is not immediately noticed when using the Accu-Chek Solo system this could potentially lead to an under-delivery of insulin.

The hazards to consumers

If insulin is under-delivered, this may lead to  adverse health events including severe hyperglycaemia or the user may experience diabetic ketoacidosis (DKA) and may require hospitalisation or intervention from a medical professional.

What consumers should do

Consumers should check whether they have an affected product and if so stop using them.  Consumers should contact Roche Diabetes Customer Care to make arrangements to receive a replacement.

For more details see https://apps.tga.gov.au/Prod/sara/arn-detail.aspx?k=RC-2022-RN-01541-1

For further information, please contact Roche Diabetes Customer Care by phone on 1800 251 516.

Details to help identify the product

Where the product was sold

New South Wales
Queensland
South Australia
Victoria
Western Australia
Australian Capital Territory

Responsible regulators

Regulators are established or appointed by government. They enforce regulations and rules.

Quote PRA number 2022/19684 when contacting a regulator about this recall.

Recall and remedy questions

Contact the Therapeutic Goods Administration if you have:

  • a question about the remedy being offered to you by the business that is responsible for managing this recall, or
  • concerns about the way the business is managing this recall.