PRA number
2020/18173
Published date
Product description
Bock Nursing Care Beds - Lifting Poles
Model Number: 423
Serial Number: 91905420-028
ARTG 209294
(Alphacare Pty Ltd - Bed, general-purpose)
Model Number: 423
Serial Number: 91905420-028
ARTG 209294
(Alphacare Pty Ltd - Bed, general-purpose)
What are the defects?
Beds with attached lifting poles, that were manufactured between April and August 2019, may have faults in the lifting poles.
What are the hazards?
Faulty-shaped lifting pole sockets can lead to slip-through of the lifting pole and may fail to sufficiently support the weight of users, resulting in a risk of injury.
What should consumers do?
Consumers with impacted stock should contact Alphacare to organise correction. Alphacare technicians will make the necessary corrections on-site.
For further information, contact Alphacare Customer Service on 1300 930 930 or via the Alphacare website at https://alphacare.com.au/contact/
For further information, contact Alphacare Customer Service on 1300 930 930 or via the Alphacare website at https://alphacare.com.au/contact/
Supplier
Where the product was sold
New South Wales
Victoria
Dates available for sale
-
Responsible regulator
Therapeutic Goods Administration is the responsible regulator for this recall.