Product description
Amico GoLift Carry Bar
Model CB-GO-LIFT-REG-1HK
Serial Numbers: CB-9541, CB-9539, CB-9538, CB-9537, CB-9536, CB-9511, CB-9488, CB-9480, CB-9479, CB-9468, CB-9467, CB-9463, CB-9462, CB-9461, CB-9460, CB-9458, CB-9457, CB-9456, CB-9455, CB-9454, CB-9443, CB-9442, CB-9439
ARTG 300589
Rehab Assist Pty Ltd - GoLift - Lift, patient transfer
Model CB-GO-LIFT-REG-1HK
Serial Numbers: CB-9541, CB-9539, CB-9538, CB-9537, CB-9536, CB-9511, CB-9488, CB-9480, CB-9479, CB-9468, CB-9467, CB-9463, CB-9462, CB-9461, CB-9460, CB-9458, CB-9457, CB-9456, CB-9455, CB-9454, CB-9443, CB-9442, CB-9439
ARTG 300589
Rehab Assist Pty Ltd - GoLift - Lift, patient transfer
See a list of details to help identify the product
Reason the product is recalled
The swivel pin may separate from the carry bar, this could result in a person falling.
A batch of carry bars were manufactured using a black oxide screw, which maybe oily, preventing the Loctite from adhering to the swivel pin assembly.
A batch of carry bars were manufactured using a black oxide screw, which maybe oily, preventing the Loctite from adhering to the swivel pin assembly.
The hazards to consumers
If the swivel pin separates from the carry bar, this could result in a person falling which may cause a serious injury.
What consumers should do
Rehab Assist is requesting consumers inspect their carry bars: if the swivel pin is exposed by 1/16” (approximately 1.5mm) and there is play in the swivel pin assembly, immediately stop using the carry bar. If consumers identify this issue they should contact Rehab Assist who will replace all carrier bars from the impacted batch.
For my information, consumers can contact Rebhab Assist Pty Ltd on 1300 364 534.
For my information, consumers can contact Rebhab Assist Pty Ltd on 1300 364 534.
Details to help identify the product
Supplier running the recall
Dates available for sale
-
Where the product was sold
Nationally
Responsible regulators
Regulators are established or appointed by government. They enforce regulations and rules.
Quote PRA number 2019/17924 when contacting a regulator about this recall.
Recall and remedy questions
Contact the Therapeutic Goods Administration if you have:
- a question about the remedy being offered to you by the business that is responsible for managing this recall, or
- concerns about the way the business is managing this recall.