Product description
CareLink (CL) software and CL System using CareLink reports
Versions: v14.1a, v3.1a and v5.2b
Product codes: MMT-7333, MMT-7350 and ACC-7333
ARTG 303062
(Medtronic Australasia Pty Ltd - Information system)
Versions: v14.1a, v3.1a and v5.2b
Product codes: MMT-7333, MMT-7350 and ACC-7333
ARTG 303062
(Medtronic Australasia Pty Ltd - Information system)
See a list of details to help identify the product.
Reason the product is recalled
Following an update to CareLink software on 29 August, 2020, consumers who generated a report from CareLink Personal or CareLink system software may have been presented with an inaccurate carbohydrate ratio setting on their CareLink report.
This only occurred under the following circumstances:
1. The carbohydrate unit of measurement in CareLink software is set to grams, but the insulin pump is set to exchanges.
2. The carbohydrate unit of measurement in CareLink software is set to exchanges, but the insulin pump is set to grams.
This only occurred under the following circumstances:
1. The carbohydrate unit of measurement in CareLink software is set to grams, but the insulin pump is set to exchanges.
2. The carbohydrate unit of measurement in CareLink software is set to exchanges, but the insulin pump is set to grams.
The hazards to consumers
If adjustments were made to the carbohydrate ratio based on the CareLink report, this could lead to either over or under delivery of insulin. This could cause low blood glucose (hypoglycemia, loss of consciousness, seizure, death), and/or high blood glucose (hyperglycemia) which can lead to serious health complications.
What consumers should do
Medtronic will contact affected users in writing. Consumers who may have adjusted their carbohydrate ratio between 29 August 2020 and 30 September 2020 using incorrect data on the CareLink report, should review their carbohydrate ratio setting with a healthcare provider to determine if any corrections are needed and should discard and re-generate the Carelink report.
On 30 September 2020, Medtronic released an automatic update to CareLink software which resolved the issue described. No further action is required to receive the corrected version of CareLink software.
For further information, please contact Medtronic Customer Service by phone on 1800 777 808.
On 30 September 2020, Medtronic released an automatic update to CareLink software which resolved the issue described. No further action is required to receive the corrected version of CareLink software.
For further information, please contact Medtronic Customer Service by phone on 1800 777 808.
Details to help identify the product.
Supplier running the recall
Where the product was sold
Nationally
International
Responsible regulators
Regulators are established or appointed by government. They enforce regulations and rules.
Quote PRA number 2020/18708 when contacting a regulator about this recall.
Recall and remedy questions
Contact the Therapeutic Goods Administration if you have:
- a question about the remedy being offered to you by the business that is responsible for managing this recall, or
- concerns about the way the business is managing this recall.