Product description

Dario Blood Glucose Tracker App
Android App Versions 4.3.0-4.3.2

ATRG 229859
(IBD Medical - Software IVDs)

An in vitro diagnostic medical device (IVD)

See a list of details to help identify the product

Reason the product is recalled

Potential issue with Android App versions of Dario Blood Glucose Tracker (4.3.0-4.3.2), regarding duplicate logging of a blood glucose level reading. The issue may cause blood glucose level readings that are transferred from the measurement screen to the 'Logbook' being replaced by an older blood glucose measurement and being displayed as a duplicate reading.

The hazards to consumers

If inaccurate blood glucose level reading from the 'Data Entry' or 'Logbook' screens are used to determine the insulin dose or make treatment decisions, this may lead to incorrect dosing decisions, which can lead to serious health complications.

What consumers should do

Update the latest Dario - Blood Glucose Tracker Android App version as follows:
Open the Google Play store
Search for 'Dario Blood Glucose Tracker' App
Press the 'Uninstall' button
Press 'OK' to confirm the question "Do you want to uninstall this app?"
Once completed, press 'Install' and verify 4.3.3 version (or above) has been installed by pressing the 'About this App' link.

For further information, contact IBD Medical on 02 8398 7520

Details to help identify the product

Supplier running the recall

Who sold the product

Google Play Store

Identifying numbers

Other identifying numbers

Android App Versions 4.3.0-4.3.2

Where the product was sold

Nationally
Online

Responsible regulators

Regulators are established or appointed by government. They enforce regulations and rules.

Quote PRA number 2019/17815 when contacting a regulator about this recall.

Recall and remedy questions

Contact the Therapeutic Goods Administration if you have:

  • a question about the remedy being offered to you by the business that is responsible for managing this recall, or
  • concerns about the way the business is managing this recall.