PRA number
2011/12260
Published date
Why the product is recalled
In an extremely weakened battery condition, the product may indicate zero volts or a low voltage reading when testing a live circuit.
What the hazards are to consumers
User may be exposed to a potential electrocution hazard.
What consumers should do
1.Stop using your clamp meter or multimeter even if you have not experienced any issues.
2.Check the model number on the front of your product and the serial number on the rear and then consult the chart above to confirm your instrument is affected.
3.Go to http://www.extech.com/instruments/recall2010.asp for immediate processing of your recall request, 24 hours a day, 7 days a week or email us at recall@extech.com. Be sure to include contact information, postal and email addresses, and model and serial numbers. We will reply to your email within one business day.
2.Check the model number on the front of your product and the serial number on the rear and then consult the chart above to confirm your instrument is affected.
3.Go to http://www.extech.com/instruments/recall2010.asp for immediate processing of your recall request, 24 hours a day, 7 days a week or email us at recall@extech.com. Be sure to include contact information, postal and email addresses, and model and serial numbers. We will reply to your email within one business day.
Supplier
Where the product was sold
Nationally
Recall advertisements and supporting documentation
Responsible regulator
Australian Competition and Consumer Commission is the responsible regulator for this recall.
