Product description
Fluke 175, 177 and 179 Digital Multimeters with serial numbers of 79000000 and below, manufactured between January 1 2001 and July 12 2001.
(Meters with serial numbers of 79000001 or higher or those units with an "R" after the serial number are not affected.
(Meters with serial numbers of 79000001 or higher or those units with an "R" after the serial number are not affected.
What are the defects?
When measuring ac voltage above 500, the display on affected meters can be slow in showing a final, accurate measurement. The final reading may not be displayed for 6 to 18 seconds. During this delay, the meter may not indicate that high voltage is present.
What are the hazards?
Electric shock.
What should consumers do?
Stop using the meter and contact VMS Laboratory on Phone (02) 9771 9300 or Facsimile (02) 9771 9600 (Attention Fluke 170 recall administrator) to arrange for repairs free of charge.