Product description
Behind-the-ear hearing aids
Multiple models and serial numbers serviced between 14 August 2019 and 5 September 2019
See attached list of models and serial numbers
ARTG 121206, 174580, 314241, 314521, 121207, 173226, 314440 & 314441
Multiple models and serial numbers serviced between 14 August 2019 and 5 September 2019
See attached list of models and serial numbers
ARTG 121206, 174580, 314241, 314521, 121207, 173226, 314440 & 314441
What are the defects?
A software error in the service and repair tool may have altered the specifications for sound amplification, presenting a risk to the user's residual hearing.
What are the hazards?
Longer exposure to mid to high level sound environment could temporarily or permanently affect or damage the user's residual hearing.
What should consumers do?
GN Hearing will contact affected users in writing. Consumers should immediately cease use of the affected products and return them to the place of purchase for a full refund.
For further information, contact GN Hearing Australia by phone on 1800 658 955.