Product description
MiniMed 640G insulin pump
An ambulatory, battery-operated, rate-programmable microinfusion pump.
Model numbers: MMT-1711 and MMT-1712
MiniMed 670G insulin pump
An insulin pump intended for continuous delivery and delivery of user activated insulin boluses.
Model numbers: MMT-1762 and MMT-1782
Only pumps with a clear pump retainer ring
ARTG 95763
(Medtronic Australasia Pty Ltd Infusion pump, insulin, ambulatory)
ARTG 308140
(Medtronic Australasia Pty Ltd - Infusion pump, insulin, ambulatory)
See a list of details to help identify the product
Reason the product is recalled
If the pump retainer ring breaks, the reservoir may become loose and may not lock into the pump.
The hazards to consumers
If the retainer ring is damaged, loose or missing, this may lead to under delivery of insulin, which may cause hyperglycemia. If a user inserts the reservoir back in while the infusion set is still connected to the body, there may be a rapid infusion of insulin, which may cause hypoglycemia or death.
What consumers should do
If the reservoir does not lock into the pump or the retainer ring is loose, damaged or missing, consumers should immediately discontinue using the insulin pump and revert to a back-up plan with the healthcare provider’s recommendations. Do not insert the reservoir back into the pump while connected.
If the reservoir properly locks in place with the retainer ring and the retainer ring is not loose, damaged, or broken, consumers may continue to use their pump until they receive a replacement pump.
In the interim, consumers should inspect the retainer ring on the pump and examine the retainer ring for damage every time the infusion set is changed.
The supplier will contact affected consumers and direct them to visit https://info.medtronicdiabetes.com/Clear_Retainer_Ring, call the Medtronic Global Help Line on 1800 777 808, or complete and return the Confirmation Form to receive a replacement pump at no charge.
Details to help identify the product
Supplier running the recall
Who sold the product
Healthcare providers
Dates available for sale
Where the product was sold
Responsible regulators
Regulators are established or appointed by government. They enforce regulations and rules.
Quote PRA number 2021/19227 when contacting a regulator about this recall.
Recall and remedy questions
Contact the Therapeutic Goods Administration if you have:
- a question about the remedy being offered to you by the business that is responsible for managing this recall, or
- concerns about the way the business is managing this recall.