Product description
MyWay Walker Chassis
Product code: 173-1600, 173-2600 and 173-3600
Serial numbers range from LEC/220568 – LEC/272111
ARTG 298435
(Sunrise Medical Pty Ltd - Holder, patient)
Product code: 173-1600, 173-2600 and 173-3600
Serial numbers range from LEC/220568 – LEC/272111
ARTG 298435
(Sunrise Medical Pty Ltd - Holder, patient)
See a list of details to help identify the product
Reason the product is recalled
The rear legs of the walker have a small cylindrical mounting feature called a boss, which the castors are screwed into. If the welding is not sufficient between the leg and the boss, the boss can come loose and the castors will detach from the rear legs of the chassis.
The hazards to consumers
If the weld fails, the boss comes loose and the castors detach from the legs of the walker, this could result in an injury and/or a fall.
What consumers should do
Sunrise Medical will contact affected consumers in writing. Consumers should send a completed acknowledgement form to Sunrise Medical to arrange for replacement legs to be fitted by an approved technician, within 4-6 months.
Consumers can continue to use their MyWay chassis after conducting a pre-use inspection to ensure that the rear castor boss is securely fixed, all castors are moving freely and all brakes and resistance features are working correctly.
For further information, contact Sunrise Medical Customer Service on 02 9678 6600.
Consumers can continue to use their MyWay chassis after conducting a pre-use inspection to ensure that the rear castor boss is securely fixed, all castors are moving freely and all brakes and resistance features are working correctly.
For further information, contact Sunrise Medical Customer Service on 02 9678 6600.
Details to help identify the product
Supplier running the recall
Who sold the product
Healthcare equipment and mobility suppliers
Where the product was sold
New South Wales
Northern Territory
Queensland
South Australia
Tasmania
Victoria
Western Australia
Responsible regulators
Regulators are established or appointed by government. They enforce regulations and rules.
Quote PRA number 2021/18933 when contacting a regulator about this recall.
Recall and remedy questions
Contact the Therapeutic Goods Administration if you have:
- a question about the remedy being offered to you by the business that is responsible for managing this recall, or
- concerns about the way the business is managing this recall.