Product description
Nokia Blood Pressure Monitor Plus (BPM+)
Model WPM03
ARTG #223033
Model WPM03
ARTG #223033
See a list of details to help identify the product
Reason the product is recalled
The blood pressure monitor did not pass applicable testing standards for accuracy for subjects with an arm circumference greater than 35cm.
The hazards to consumers
The user may be unable to obtain an accurate reading of the blood pressure, including systolic and diastolic readings, if the subject's arm circumference is greater than 35cm.
What consumers should do
Users should have been contacted by an email in June 2018 and an application update in July 2018, which requires them to select one of the following options:
1. Keep their Nokia BPM+, which requires that the user acknowledges that they have been notified
2. Return their Nokia BPM+ and receive a refund. If this option is selected a link to the return form is provided.
For more information, contact Emergo Customer Care on 02 9006 1662.
1. Keep their Nokia BPM+, which requires that the user acknowledges that they have been notified
2. Return their Nokia BPM+ and receive a refund. If this option is selected a link to the return form is provided.
For more information, contact Emergo Customer Care on 02 9006 1662.
Details to help identify the product
Supplier running the recall
Where the product was sold
Nationally
Responsible regulators
Regulators are established or appointed by government. They enforce regulations and rules.
Quote PRA number 2018/17121 when contacting a regulator about this recall.
Recall and remedy questions
Contact the Therapeutic Goods Administration if you have:
- a question about the remedy being offered to you by the business that is responsible for managing this recall, or
- concerns about the way the business is managing this recall.