PRA number
2019/17924
Published date
Product description
Amico GoLift Carry Bar
Model CB-GO-LIFT-REG-1HK
Serial Numbers: CB-9541, CB-9539, CB-9538, CB-9537, CB-9536, CB-9511, CB-9488, CB-9480, CB-9479, CB-9468, CB-9467, CB-9463, CB-9462, CB-9461, CB-9460, CB-9458, CB-9457, CB-9456, CB-9455, CB-9454, CB-9443, CB-9442, CB-9439
ARTG 300589
Rehab Assist Pty Ltd - GoLift - Lift, patient transfer
Model CB-GO-LIFT-REG-1HK
Serial Numbers: CB-9541, CB-9539, CB-9538, CB-9537, CB-9536, CB-9511, CB-9488, CB-9480, CB-9479, CB-9468, CB-9467, CB-9463, CB-9462, CB-9461, CB-9460, CB-9458, CB-9457, CB-9456, CB-9455, CB-9454, CB-9443, CB-9442, CB-9439
ARTG 300589
Rehab Assist Pty Ltd - GoLift - Lift, patient transfer
Reasons the product is recalled
The swivel pin may separate from the carry bar, this could result in a person falling.
A batch of carry bars were manufactured using a black oxide screw, which maybe oily, preventing the Loctite from adhering to the swivel pin assembly.
A batch of carry bars were manufactured using a black oxide screw, which maybe oily, preventing the Loctite from adhering to the swivel pin assembly.
The hazards to consumers
If the swivel pin separates from the carry bar, this could result in a person falling which may cause a serious injury.
What consumers should do
Rehab Assist is requesting consumers inspect their carry bars: if the swivel pin is exposed by 1/16” (approximately 1.5mm) and there is play in the swivel pin assembly, immediately stop using the carry bar. If consumers identify this issue they should contact Rehab Assist who will replace all carrier bars from the impacted batch.
For my information, consumers can contact Rebhab Assist Pty Ltd on 1300 364 534.
For my information, consumers can contact Rebhab Assist Pty Ltd on 1300 364 534.
Responsible regulator
Therapeutic Goods Administration is the responsible regulator for this recall.