Product description
MEDUCORE Easy Automatic External Defibrillator (AED)
See a list of details to help identify the product
Reason the product is recalled
During safety checks, it was discovered that the time allotted for charging the capacitor was insufficient for a full charge to be made.
The hazards to consumers
Potentially life-threatening, or could cause a serious risk to health (Class I)
What consumers should do
Please do not contact the sponsor for further information unless you believe that you have the goods under recall and have not received a recall letter. Refer to the following website for further details:
http://www.tga.gov.au/safety/alerts-device-meducore-defibrillator-111006.htm
http://www.tga.gov.au/safety/alerts-device-meducore-defibrillator-111006.htm
Details to help identify the product
Supplier running the recall
Who sold the product
Weinmann Australia Pty Ltd, Ferno Australia Pty Ltd.
Identifying product features
Ferno Australia Pty Ltd is seeking to locate Meducore Easy Automatic External Defibrillators bearing the following serial numbers:
1226, 3163, 3164, 3165, 3167, 3171, 3172, 3176, 3177, 3178, 3182, 3183, 3184, 3185, 3186, 3188, 3189, 3192, 3193, 3199, 3200, 3208
Identifying numbers
Other identifying numbers
Ferno Australia Pty Ltd is seeking to locate Meducore Easy Automatic External Defibrillators bearing the following serial numbers:
1226, 3163, 3164, 3165, 3167, 3171, 3172, 3176, 3177, 3178, 3182, 3183, 3184, 3185, 3186, 3188, 3189, 3192, 3193, 3199, 3200, 3208
1226, 3163, 3164, 3165, 3167, 3171, 3172, 3176, 3177, 3178, 3182, 3183, 3184, 3185, 3186, 3188, 3189, 3192, 3193, 3199, 3200, 3208
Where the product was sold
Australian Capital Territory
New South Wales
Queensland
Victoria
Western Australia
Responsible regulators
Regulators are established or appointed by government. They enforce regulations and rules.
Quote PRA number 2011/12610 when contacting a regulator about this recall.
Recall and remedy questions
Contact the Therapeutic Goods Administration if you have:
- a question about the remedy being offered to you by the business that is responsible for managing this recall, or
- concerns about the way the business is managing this recall.