Arjo Australia Pty Ltd— Nimbus 4 and Nimbus Professional Mattresses

PRA number

2021/19012

Published date

Product description

Nimbus 4 and Nimbus Professional Mattresses

Serial numbers: 1308000048 to 2000027834

ARTG 289068
(Arjo Australia Pty Ltd - Pressure alleviation pad)
Identifying features

Other identifying numbers

Model numbers: 650001ADV, 650001DAR, 650001P, 650201DAR, 651001, 651001ADV, 651001DAR, 651001DARW, 651001P, 651201DAR, 651201P

Reasons the product is recalled

Nimbus 4 and Nimbus Professional mattresses may over-inflate, due to an inner mattress component failure. This cannot be detected until the product is disassembled.

The hazards to consumers

If the mattress over-inflates, patients could fall from the mattress and this could result in serious injury.

What consumers should do

Affected consumers will be contacted in writing. Consumers should follow the instructions provided in the customer letter and the instructional video, to puncture the grommet membrane so that excess air may escape. Removing the membrane will not affect the mattress functionality in any way.

For more information, please contact Arjo Australia by phone on 1800 072 040.

Where the product was sold

Australian Capital Territory
New South Wales
Northern Territory
Queensland
South Australia
Victoria
Western Australia

Dates available for sale

-

Responsible regulator

Therapeutic Goods Administration is the responsible regulator for this recall.

Product category