Product description
Fluke 37X Series Clamp Meters
See a list of details to help identify the product
Reason the product is recalled
A screw connecting the printed circuit assembly to the test lead input jack can work itself loose and result in intermittent voltage readings, including a ‘0’ reading when voltage is present.
The hazards to consumers
If the clamp meter is not reading correctly, users may be exposed to a potential electrocution hazard.
What consumers should do
1. Stop using your affected 37X series clamp meter(s) even if you have not experienced problems.
2. Verify that your clamp meter is affected using the serial numbers provided in the table.
3. If the product falls within the serial number range, return the unit to Celemetrix for repair.
2. Verify that your clamp meter is affected using the serial numbers provided in the table.
3. If the product falls within the serial number range, return the unit to Celemetrix for repair.
Details to help identify the product
Supplier running the recall
Who sold the product
Electrical trade and retail outlets nationally
Identifying product features
Models affected:
Fluke-376
Fluke-375
Fluke-374
Fluke-373
Please see the attached list for further information on manufacture dates and serial numbers.
Identifying numbers
Other identifying numbers
Models affected:
Fluke-376
Fluke-375
Fluke-374
Fluke-373
Please see the attached list for further information on manufacture dates and serial numbers.
Fluke-376
Fluke-375
Fluke-374
Fluke-373
Please see the attached list for further information on manufacture dates and serial numbers.
Where the product was sold
Nationally
Responsible regulators
Regulators are established or appointed by government. They enforce regulations and rules.
Quote PRA number 2013/13457 when contacting a regulator about this recall.
Recall and remedy questions
Contact the Australian Competition and Consumer Commission if you have:
- a question about the remedy being offered to you by the business that is responsible for managing this recall, or
- concerns about the way the business is managing this recall.