Product description
Fluke 333, 334, 335, 336 and 337 Digital Clamp Meters manufactured between 3 March 2008 and 27 February 2009.
See a list of details to help identify the product
Reason the product is recalled
A defect has the potential, over time, to cause the meter to display a low or no voltage reading on a circuit energised with a hazardous voltage. In this situation it is imperative that the practice of testing the meter on a known live circuit before and after checking the circuit under test is maintained. Should the meter not pass the "known live circuit test" its use should be discontinued immediately, identified as faulty and removed from service.
The hazards to consumers
Electrocution due to incorrect reading
What consumers should do
Cease using and immediately return suspect products to the place of purchase to be replaced free of charge. For further information visit www.fluke.com.au or phone (02) 8850 3333.
Details to help identify the product
Supplier running the recall
Identifying product features
Fluke 333, 334, 335, 336 and 337 Digital Clamp Meters
Identifying numbers
Other identifying numbers
Fluke 333, 334, 335, 336 and 337 Digital Clamp Meters
Where the product was sold
Nationally
Responsible regulators
Regulators are established or appointed by government. They enforce regulations and rules.
Quote PRA number 2009/10778 when contacting a regulator about this recall.
Recall and remedy questions
Contact the Australian Competition and Consumer Commission if you have:
- a question about the remedy being offered to you by the business that is responsible for managing this recall, or
- concerns about the way the business is managing this recall.