Product description

Serial Numbers Between 70521601 And 74165430. If The T2 Has A Serial Number Lower Than 70521601 And It Does Not Have An "r" Stamped After The Serial Number, The T2 Was Not Repaired In The November 1998 Recall And Must Be Returned To Fluke For Repair.

See a list of details to help identify the product

Reason the product is recalled

The Positive Battery Contact Can Corrode Due To Exposure To Vibration And The Instrument Can Operate Intermittently. When The Malfunction Occurs, The Tester May Not Indicate That Voltage Is Present Placing The User In A Potentially Hazardous Situation.

The hazards to consumers

Electric Shock

What consumers should do

Either Return With A Safety Notice Return Form Available By Faxing Fluke Australia (02) 8850 3300; Or Request A Free Field Repair Kit By Telephoning (02) 8850 3333 Or Faxing Fluke Australia.

Details to help identify the product

Supplier running the recall

Fluke Australia Pty Ltd

Responsible regulators

Regulators are established or appointed by government. They enforce regulations and rules.

Quote PRA number 1999/4103 when contacting a regulator about this recall.

Recall and remedy questions

Contact the Australian Competition and Consumer Commission if you have:

  • a question about the remedy being offered to you by the business that is responsible for managing this recall, or
  • concerns about the way the business is managing this recall.